A correlation study of eye lens dose and personal dose equivalent for interventional cardiologists

Research output: Contribution to journalArticleScientificpeer-review

Details

Original languageEnglish
Article numbernct180
Pages (from-to)561-569
Number of pages9
JournalRadiation Protection Dosimetry
Volume157
Issue number4
DOIs
Publication statusPublished - Dec 2013
Publication typeA1 Journal article-refereed

Abstract

This paper presents the dosimetry part of the European ELDO project, funded by the DoReMi Network of Excellence, in which a method was developed to estimate cumulative eye lens doses for past practices based on personal dose equivalent values, Hp(10), measured above the lead apron at several positions at the collar, chest and waist levels. Measurement campaigns on anthropomorphic phantoms were carried out in typical interventional settings considering different tube projections and configurations, beam energies and filtration, operator positions and access routes and using both mono-tube and biplane X-ray systems. Measurements showed that eye lens dose correlates best with Hp(10) measured on the left side of the phantom at the level of the collar, although this correlation implicates high spreads (41 %). Nonetheless, for retrospective dose assessment, Hp(10) records are often the only option for eye dose estimates and the typically used chest left whole-body dose measurement remains useful.