Accuracy of linear intraoral measurements using cone beam CT and multidetector CT: A tale of two CTs

Research output: Contribution to journalArticleScientificpeer-review

Details

Original languageEnglish
Pages (from-to)637-644
Number of pages8
JournalDentomaxillofacial Radiology
Volume41
Issue number8
DOIs
Publication statusPublished - 1 Dec 2012
Publication typeA1 Journal article-refereed

Abstract

The aim was to compare the accuracy of linear bone measurements of cone beam CT (CBCT) with multidetector CT (MDCT) and validate intraoral soft-tissue measurements in CBCT. Methods: Comparable views of CBCT and MDCT were obtained from eight intact cadaveric heads. The anatomical positions of the gingival margin and the buccal alveolar bone ridge were determined. Image measurements (CBCT/MDCT) were performed upon multiplanar reformatted data sets and compared with the anatomical measurements; the number of non-assessable sites (NASs) was evaluated. Results: Radiological measurements were accurate with a mean difference from anatomical measurements of 0.14mm (CBCT) and 0.23mm (MDCT). These differences were statistically not significant, but the limits of agreement for bone measurements were broader in MDCT (21.35 mm; 1.82 mm) than in CBCT (-0.93 mm; 1.21 mm). The limits of agreement for soft tissue measurements in CBCT were smaller (-0.77 mm; 1.07 mm), indicating a slightly higher accuracy. More NASs occurred in MDCT (14.5%) than in CBCT (8.3%).

Keywords

  • Accuracy, CBCT, Computed tomography, Image quality, Soft tissue