Lead field theory provides a powerful tool for designing microelectrode array impedance measurements for biological cell detection and observation

Research output: Contribution to journalArticleScientificpeer-review

Standard

Lead field theory provides a powerful tool for designing microelectrode array impedance measurements for biological cell detection and observation. / Böttrich, Marcel; Tanskanen, Jarno M.A.; Hyttinen, Jari A.K.

In: BioMedical Engineering Online, Vol. 16, No. 1, 85, 26.06.2017.

Research output: Contribution to journalArticleScientificpeer-review

Harvard

APA

Vancouver

Author

Bibtex - Download

@article{3494179d42f04abdb24944de60f09a72,
title = "Lead field theory provides a powerful tool for designing microelectrode array impedance measurements for biological cell detection and observation",
abstract = "Background: Our aim is to introduce a method to enhance the design process of microelectrode array (MEA) based electric bioimpedance measurement systems for improved detection and viability assessment of living cells and tissues. We propose the application of electromagnetic lead field theory and reciprocity for MEA design and measurement result interpretation. Further, we simulated impedance spectroscopy (IS) with two- and four-electrode setups and a biological cell to illustrate the tool in the assessment of the capabilities of given MEA electrode constellations for detecting cells on or in the vicinity of the microelectrodes. Results: The results show the power of the lead field theory in electromagnetic simulations of cell-microelectrode systems depicting the fundamental differences of two- and four-electrode IS measurement configurations to detect cells. Accordingly, the use in MEA system design is demonstrated by assessing the differences between the two- and four-electrode IS configurations. Further, our results show how cells affect the lead fields in these MEA system, and how we can utilize the differences of the two- and four-electrode setups in cell detection. The COMSOL simulator model is provided freely in public domain as open source. Conclusions: Lead field theory can be successfully applied in MEA design for the IS based assessment of biological cells providing the necessary visualization and insight for MEA design. The proposed method is expected to enhance the design and usability of automated cell and tissue manipulation systems required for bioreactors, which are intended for the automated production of cell and tissue grafts for medical purposes. MEA systems are also intended for toxicology to assess the effects of chemicals on living cells. Our results demonstrate that lead field concept is expected to enhance also the development of such methods and devices.",
keywords = "Bioimpedance, Biological cells, Finite element analysis, Impedance spectroscopy, Microelectrodes, Sensor arrays",
author = "Marcel B{\"o}ttrich and Tanskanen, {Jarno M.A.} and Hyttinen, {Jari A.K.}",
year = "2017",
month = "6",
day = "26",
doi = "10.1186/s12938-017-0372-5",
language = "English",
volume = "16",
journal = "BioMedical Engineering Online",
issn = "1475-925X",
publisher = "Springer Verlag",
number = "1",

}

RIS (suitable for import to EndNote) - Download

TY - JOUR

T1 - Lead field theory provides a powerful tool for designing microelectrode array impedance measurements for biological cell detection and observation

AU - Böttrich, Marcel

AU - Tanskanen, Jarno M.A.

AU - Hyttinen, Jari A.K.

PY - 2017/6/26

Y1 - 2017/6/26

N2 - Background: Our aim is to introduce a method to enhance the design process of microelectrode array (MEA) based electric bioimpedance measurement systems for improved detection and viability assessment of living cells and tissues. We propose the application of electromagnetic lead field theory and reciprocity for MEA design and measurement result interpretation. Further, we simulated impedance spectroscopy (IS) with two- and four-electrode setups and a biological cell to illustrate the tool in the assessment of the capabilities of given MEA electrode constellations for detecting cells on or in the vicinity of the microelectrodes. Results: The results show the power of the lead field theory in electromagnetic simulations of cell-microelectrode systems depicting the fundamental differences of two- and four-electrode IS measurement configurations to detect cells. Accordingly, the use in MEA system design is demonstrated by assessing the differences between the two- and four-electrode IS configurations. Further, our results show how cells affect the lead fields in these MEA system, and how we can utilize the differences of the two- and four-electrode setups in cell detection. The COMSOL simulator model is provided freely in public domain as open source. Conclusions: Lead field theory can be successfully applied in MEA design for the IS based assessment of biological cells providing the necessary visualization and insight for MEA design. The proposed method is expected to enhance the design and usability of automated cell and tissue manipulation systems required for bioreactors, which are intended for the automated production of cell and tissue grafts for medical purposes. MEA systems are also intended for toxicology to assess the effects of chemicals on living cells. Our results demonstrate that lead field concept is expected to enhance also the development of such methods and devices.

AB - Background: Our aim is to introduce a method to enhance the design process of microelectrode array (MEA) based electric bioimpedance measurement systems for improved detection and viability assessment of living cells and tissues. We propose the application of electromagnetic lead field theory and reciprocity for MEA design and measurement result interpretation. Further, we simulated impedance spectroscopy (IS) with two- and four-electrode setups and a biological cell to illustrate the tool in the assessment of the capabilities of given MEA electrode constellations for detecting cells on or in the vicinity of the microelectrodes. Results: The results show the power of the lead field theory in electromagnetic simulations of cell-microelectrode systems depicting the fundamental differences of two- and four-electrode IS measurement configurations to detect cells. Accordingly, the use in MEA system design is demonstrated by assessing the differences between the two- and four-electrode IS configurations. Further, our results show how cells affect the lead fields in these MEA system, and how we can utilize the differences of the two- and four-electrode setups in cell detection. The COMSOL simulator model is provided freely in public domain as open source. Conclusions: Lead field theory can be successfully applied in MEA design for the IS based assessment of biological cells providing the necessary visualization and insight for MEA design. The proposed method is expected to enhance the design and usability of automated cell and tissue manipulation systems required for bioreactors, which are intended for the automated production of cell and tissue grafts for medical purposes. MEA systems are also intended for toxicology to assess the effects of chemicals on living cells. Our results demonstrate that lead field concept is expected to enhance also the development of such methods and devices.

KW - Bioimpedance

KW - Biological cells

KW - Finite element analysis

KW - Impedance spectroscopy

KW - Microelectrodes

KW - Sensor arrays

U2 - 10.1186/s12938-017-0372-5

DO - 10.1186/s12938-017-0372-5

M3 - Article

VL - 16

JO - BioMedical Engineering Online

JF - BioMedical Engineering Online

SN - 1475-925X

IS - 1

M1 - 85

ER -