Tampere University of Technology

TUTCRIS Research Portal

Scanner abdominal: Étude comparative de l'exposition patient en routine clinique sur des appareils avec et sans reconstruction itérative

Research output: Contribution to journalArticleScientificpeer-review

Details

Original languageFrench
Pages (from-to)35-41
Number of pages7
JournalRadioprotection
Volume49
Issue number1
DOIs
Publication statusPublished - 2014
Publication typeA1 Journal article-refereed

Abstract

Objective: compare the dose delivered to patients and image quality in clinical routine to perform an abdominal CT scan with no iterative reconstruction techniques (IR) relative to an examination conducted on a scanner with IR. Materials and methods: this is a retrospective study of 30 patients who underwent two abdominal examinations: one on a 40-slice scanner (TDM40) without IR and another one on a 256-slice scanner with IR (TDM256). The patients, on medical follow-up for a chronic abdominal disease, had an exam on each scanner using the same protocol comprising an abdominopelvic time portal phase. The length of acquisition, the effective dose and the dose length product (DLP) as well as quantitative and qualitative assessments of the image were compared. Results: the average effective dose per examination was 17.3 mSv with the TDM40 (PDL: 1019 mGy.cm) against 11.1 mSv with the TDM256 (PDL: 654 mGy.cm), hence a reduction of 35.8% (p <0.001). The length of acquisition and quantification were comparable in both groups. The qualitative assessment was slightly higher on the TDM40 but no examination was considered suboptimal. Conclusion: using a scanner equipped with IR significantly reduces the effective dose while maintaining image quality.