Designing, Manufacturing and Testing of a Piezoelectric Polymer Film In-Sole Sensor for Plantar Pressure Distribution Measurements

Tutkimustuotosvertaisarvioitu

Yksityiskohdat

AlkuperäiskieliEnglanti
Sivut6798-6805
Sivumäärä8
JulkaisuIEEE Sensors Journal
Vuosikerta17
Numero20
DOI - pysyväislinkit
TilaJulkaistu - 15 lokakuuta 2017
OKM-julkaisutyyppiA1 Alkuperäisartikkeli

Tiivistelmä

We have designed, manufactured, and tested a prototype of an in-sole plantar pressure sensor. The sensor is made of a piezoelectric polymer film (polyvinylidenefluoride, PVDF) with evaporated copper electrodes. The size and the location of measuring sites are carefully considered to produce a light-weighted, practical, and inexpensive matrix sensor for measurement of plantar pressure distribution. The sensor consists of eight measurement locations: hallux, first metatarsal head (under both sesamoid bones), metatarsal heads 2-5 and heel. The developed in-sole sensor was evaluated through piezoelectric sensitivity measurements and test measurements with subjects (used as a platform sensor). Average peak-to-peak pressures from 58 to 486 kPa were obtained in plantar pressure measurements. The in-sole sensor can be used to prevent pressure ulcers. Also other application areas for the sensor are available, e.g., rehabilitation and sports. The developed sensor enables unobtrusive and long-term plantar pressure measurements. In future, an option for manufacturing the PVDF in-soles may be printing of desired electrode pattern on PVDF material, that would allow fast production, and if needed, the production of even individual in-soles.