Flow rate-independent electrical aerosol sensor

Tutkimustuotosvertaisarvioitu

Yksityiskohdat

AlkuperäiskieliEnglanti
Sivut1283-1292
JulkaisuAerosol Science and Technology
Vuosikerta52
Numero11
Varhainen verkossa julkaisun päivämäärä2018
DOI - pysyväislinkit
TilaJulkaistu - 2018
OKM-julkaisutyyppiA1 Alkuperäisartikkeli

Tiivistelmä

A new diffusion charging-based aerosol instrument design is presented, intended to be a starting point for a low-cost particle concentration sensor. The aim in the design is to minimize instrument response dependence on the sample flow rate. The operation principle, response functions of the components, and performance of a prototype instrument are reported. Based on the performance evaluation, the instrument response remained constant within ±15% over a wide sample flow rate range of 3 to 10 lpm. While in the design some sensitivity is sacrificed to minimize flow rate dependency, the detection limit is still sufficient for a practical sensor application.